MDR and IVDR classification
Updated on November 15, 2026
In brief
Annex VIII of the MDR sets out 22 classification rules, Annex VIII of the IVDR 7. The tool below asks the questions that decide between them, shows at each step the official text the question rests on, then gives the indicative class, the rule or rules involved and the Normiq roadmap for that class in the European Union, Switzerland or both. Classification remains the manufacturer's responsibility: the tool does not replace it, it documents it.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify a device
Answer “I don’t know” when unsure: the tool does not guess, it then shows every class that remains possible.
Which regulation applies to your device?
The 29 rules, one page each
Each page gives the official text of the rule in your language, an explanation, the examples from the MDCG guidance and the Normiq roadmap for each class it assigns.
Medical device — Regulation (EU) 2017/745
Annex VIII, 4. NON-INVASIVE DEVICES
- Rule 1: Non-invasive devices: the default ruleClass I
- Rule 2: Channelling or storing blood, liquids, tissues or gasesClass I · Class IIa · Class IIb
- Rule 3: Modifying the composition of blood, tissues or liquidsClass IIa · Class IIb · Class III
- Rule 4: Contact with injured skin or mucous membraneClass I · Class IIa · Class IIb
Annex VIII, 5. INVASIVE DEVICES
- Rule 5: Devices invasive with respect to body orificesClass I · Class IIa · Class IIb
- Rule 6: Surgically invasive devices for transient useClass I · Class IIa · Class IIb · Class III
- Rule 7: Surgically invasive devices for short-term useClass IIa · Class IIb · Class III
- Rule 8: Implantable and long-term surgically invasive devicesClass IIa · Class IIb · Class III
Annex VIII, 6. ACTIVE DEVICES
- Rule 9: Active therapeutic devicesClass IIa · Class IIb · Class III
- Rule 10: Active devices for diagnosis and monitoringClass I · Class IIa · Class IIb
- Rule 11: SoftwareClass I · Class IIa · Class IIb · Class III
- Rule 12: Administering or removing medicinal products, body liquids or substancesClass IIa · Class IIb
- Rule 13: All other active devicesClass I
Annex VIII, 7. SPECIAL RULES
- Rule 14: Devices incorporating a medicinal substanceClass III
- Rule 15: Contraception and prevention of sexually transmitted diseasesClass IIb · Class III
- Rule 16: Disinfection, sterilisation and contact lens careClass IIa · Class IIb
- Rule 17: Recording X-ray diagnostic imagesClass IIa
- Rule 18: Tissues or cells of human or animal originClass III
- Rule 19: NanomaterialsClass IIa · Class IIb · Class III
- Rule 20: Administering medicinal products by inhalationClass IIa · Class IIb
- Rule 21: Devices composed of substancesClass IIa · Class IIb · Class III
- Rule 22: Active therapeutic devices with an integrated diagnostic functionClass III
In vitro diagnostic medical device — Regulation (EU) 2017/746
Annex VIII, 2. CLASSIFICATION RULES
- Rule 1: Transmissible agents: donations, propagation, infectious loadClass D
- Rule 2: Blood grouping and tissue typingClass C · Class D
- Rule 3: The class C listClass C
- Rule 4: Self-testing and near-patient testingClass B · Class C
- Rule 5: Laboratory products, instruments and specimen receptaclesClass A
- Rule 6: Devices not covered by the other rulesClass B
- Rule 7: Controls without an assigned valueClass B
Implementing rules
Before the rules, Annex VIII sets out how to apply them: the intended purpose of the device governs, and where several rules apply, the strictest prevails.
MDR: the 7 implementing points
- Annex VIII, Section 3.1
Application of the classification rules shall be governed by the intended purpose of the devices.
- Annex VIII, Section 3.2
If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices. Accessories for a medical device shall be classified in their own right separately from the device with which they are used.
- Annex VIII, Section 3.3
Software, which drives a device or influences the use of a device, shall fall within the same class as the device.
- If the software is independent of any other device, it shall be classified in its own right.
- Annex VIII, Section 3.4
If the device is not intended to be used solely or principally in a specific part of the body, it shall be considered and classified on the basis of the most critical specified use.
- Annex VIII, Section 3.5
If several rules, or if, within the same rule, several sub-rules, apply to the same device based on the device's intended purpose, the strictest rule and sub-rule resulting in the higher classification shall apply.
- Annex VIII, Section 3.6
In calculating the duration referred to in Section 1, continuous use shall mean:
- the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and
- the accumulated use of a device that is intended by the manufacturer to be replaced immediately with another of the same type.
- Annex VIII, Section 3.7
A device is considered to allow direct diagnosis when it provides the diagnosis of the disease or condition in question by itself or when it provides decisive information for the diagnosis.
IVDR: the 10 implementing points
- Annex VIII, Section 1.1
Application of the classification rules shall be governed by the intended purpose of the devices.
- Annex VIII, Section 1.2
If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices.
- Annex VIII, Section 1.3
Accessories for an in vitro diagnostic medical device shall be classified in their own right separately from the device with which they are used.
- Annex VIII, Section 1.4
Software, which drives a device or influences the use of a device, shall fall within the same class as the device.
- If the software is independent of any other device, it shall be classified in its own right.
- Annex VIII, Section 1.5
Calibrators intended to be used with a device shall be classified in the same class as the device.
- Annex VIII, Section 1.6
Control materials with quantitative or qualitative assigned values intended for one specific analyte or multiple analytes shall be classified in the same class as the device.
- Annex VIII, Section 1.7
The manufacturer shall take into consideration all classification and implementation rules in order to establish the proper classification for the device.
- Annex VIII, Section 1.8
Where a manufacturer states multiple intended purposes for a device, and as a result the device falls into more than one class, it shall be classified in the higher class.
- Annex VIII, Section 1.9
If several classification rules apply to the same device, the rule resulting in the higher classification shall apply.
- Annex VIII, Section 1.10
Each of the classification rules shall apply to first line assays, confirmatory assays and supplemental assays.
Who classifies the device
Classification is the manufacturer's task (MDR Art. 51; IVDR Art. 47), recorded with the justification of the rules applied in its technical documentation. The class the tool gives is indicative: it repeats what the cited rules assign to your answers, never more. A dispute with the notified body over the application of Annex VIII is decided by the competent authority.
MDR
MDR, Article 51, paragraph 1
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. Classification shall be carried out in accordance with Annex VIII.
Annex II, Section 1.1, f)
the risk class of the device and the justification for the classification rule(s) applied in accordance with Annex VIII;
IVDR
IVDR, Article 47, paragraph 1
Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. Classification shall be carried out in accordance with Annex VIII.
Annex II, Section 1.1, f)
the risk class of the device and the justification for the classification rule(s) applied in accordance with Annex VIII;
In Switzerland
The MedDO and the IvDO do not rewrite the rules: they refer to Annex VIII of the EU regulation. The class is therefore the same on both markets.
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.