IVDR · Annex VIII, Section 2.7 · CLASSIFICATION RULES
IVDR rule 7: Controls without an assigned value
Updated on November 15, 2026
Official text
Devices which are controls without a quantitative or qualitative assigned value are classified as class B.
IVDR, Annex VIII, Section 2.7, consolidated version of January 10, 2025.
Short link: normiq.app/en/ivdr/annexe/viii/7
In plain words
A control without a quantitative or qualitative assigned value is in class B. A control material with quantitative or qualitative assigned values intended for one or more analytes instead takes the class of the device (Section 1.6).
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 7 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).
Class B
Unassigned control sera.
MDCG 2020-16 rev.5, 4, Rule 7, p. 45
Non-assay specific control plasma for use in coagulation.
MDCG 2020-16 rev.5, 4, Rule 7, p. 45
Classes and Normiq roadmaps
Rule 7 assigns: Class B. For each, the Normiq roadmap on every market where Annex VIII applies.
The same rule in Switzerland
Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
Classify your device
The tool asks the questions of the 7 rules of the IVDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool