MDR · Annex VIII, Section 7.9 · SPECIAL RULES
MDR rule 22: Active therapeutic devices with an integrated diagnostic function
Updated on November 15, 2026
Official text
Active therapeutic devices with an integrated or incorporated diagnostic function which significantly determines the patient management by the device, such as closed loop systems or automated external defibrillators, are classified as class III.
MDR, Annex VIII, Section 7.9, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/22
In plain words
An active therapeutic device with an integrated or incorporated diagnostic function which significantly determines the patient management by the device, such as a closed loop system or an automated external defibrillator, is in class III.
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 4
‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.
- Software shall also be deemed to be an active device;
Examples from the MDCG guidance
The guidance lists these examples under rule 22 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class III
Automated closed loop insulin delivery system
MDCG 2021-24 Rev.1, 4.2, Rule 22, p. 59
Semiautomatic external defibrillators
MDCG 2021-24 Rev.1, 4.2, Rule 22, p. 59
Classes and Normiq roadmaps
Rule 22 assigns: Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool