IVDR · Annex VIII, Section 2.1 · CLASSIFICATION RULES
IVDR rule 1: Transmissible agents: donations, propagation, infectious load
Updated on November 15, 2026
Official text
Devices intended to be used for the following purposes are classified as class D:
- detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration;
- detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation;
- determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.
IVDR, Annex VIII, Section 2.1, consolidated version of January 10, 2025.
Short link: normiq.app/en/ivdr/annexe/viii/1
In plain words
Class D covers devices intended to detect the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or their derivatives, to assess their suitability for transfusion, transplantation or cell administration; to detect the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; or to determine the infectious load of a life-threatening disease where monitoring is critical in patient management.
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 1 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).
Class D
A device intended to screen blood and tissue donations for syphilis would fall under class D according to rule 1.
MDCG 2020-16 rev.5, 3.1 Intended purpose, p. 5
Devices intended to be used for determining the infectious load of: Human Immunodeficiency Virus.
MDCG 2020-16 rev.5, 4, Rule 1 third indent, p. 13
Classes and Normiq roadmaps
Rule 1 assigns: Class D. For each, the Normiq roadmap on every market where Annex VIII applies.
The same rule in Switzerland
Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
Classify your device
The tool asks the questions of the 7 rules of the IVDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool