MDR · Annex VIII, Section 4.2 · NON-INVASIVE DEVICES
MDR rule 2: Channelling or storing blood, liquids, tissues or gases
Updated on November 15, 2026
Official text
All non-invasive devices intended for channelling or storing blood, body liquids, cells or tissues, liquids or gases for the purpose of eventual infusion, administration or introduction into the body are classified as class IIa:
- if they may be connected to a class IIa, class IIb or class III active device; or
- if they are intended for use for channelling or storing blood or other body liquids or for storing organs, parts of organs or body cells and tissues, except for blood bags; blood bags are classified as class IIb.
In all other cases, such devices are classified as class I.
MDR, Annex VIII, Section 4.2, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/2
In plain words
The rule covers non-invasive devices that channel or store blood, body liquids, cells or tissues, liquids or gases for eventual infusion, administration or introduction into the body. They are in class IIa if they may be connected to an active device of class IIa, IIb or III, or if they channel or store blood or other body liquids, or store organs, cells or tissues; blood bags are in class IIb, and all others in class I.
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 2 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class I
Non-invasive devices that provide a simple channelling function, with gravity providing the force to transport the liquid, e.g. administration sets for infusion
MDCG 2021-24 Rev.1, 4.2, Rule 2, p. 30
Class IIa
Syringes for infusion pumps
MDCG 2021-24 Rev.1, 4.2, Rule 2, p. 29
Class IIb
Blood bags without a substance which, if used separately, can be considered to be a medicinal product
MDCG 2021-24 Rev.1, 4.2, Rule 2, p. 30
Classes and Normiq roadmaps
Rule 2 assigns: Class I, Class IIa, Class IIb. For each, the Normiq roadmap on every market where Annex VIII applies.
Class I
Class IIa
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool