MDR · Annex VIII, Section 7.2 · SPECIAL RULES
MDR rule 15: Contraception and prevention of sexually transmitted diseases
Updated on November 15, 2026
Official text
All devices used for contraception or prevention of the transmission of sexually transmitted diseases are classified as class IIb, unless they are implantable or long term invasive devices, in which case they are classified as class III.
MDR, Annex VIII, Section 7.2, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/15
In plain words
A device used for contraception or prevention of the transmission of sexually transmitted diseases is in class IIb, or class III if it is implantable or long-term invasive.
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 6
‘invasive device’ means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body;
- MDR, Annex VIII, Section 2.1
‘Body orifice’ means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.
- MDR, Annex VIII, Section 2.2
‘Surgically invasive device’ means:
- an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and
- a device which produces penetration other than through a body orifice.
- MDR, Article 2, point 5
‘implantable device’ means any device, including those that are partially or wholly absorbed, which is intended:
- to be totally introduced into the human body, or
- to replace an epithelial surface or the surface of the eye,
- by clinical intervention and which is intended to remain in place after the procedure.
- Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device;
- MDR, Annex VIII, Section 1.1
‘Transient’ means normally intended for continuous use for less than 60 minutes.
- MDR, Annex VIII, Section 1.2
‘Short term’ means normally intended for continuous use for between 60 minutes and 30 days.
- MDR, Annex VIII, Section 1.3
‘Long term’ means normally intended for continuous use for more than 30 days.
- MDR, Annex VIII, Section 3.6
In calculating the duration referred to in Section 1, continuous use shall mean:
- the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and
- the accumulated use of a device that is intended by the manufacturer to be replaced immediately with another of the same type.
Examples from the MDCG guidance
The guidance lists these examples under rule 15 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class IIb
Condoms and femidoms (internal condoms)
MDCG 2021-24 Rev.1, 4.2, Rule 15, p. 51
Contraceptive diaphragms
MDCG 2021-24 Rev.1, 4.2, Rule 15, p. 51
Fertility monitors and medical device software intended to be used in contraception (e.g. by using the basal body temperature)
MDCG 2021-24 Rev.1, 4.2, Rule 15, p. 51
Class III
Tubal ligation devices (e.g. clips or rings)
MDCG 2021-24 Rev.1, 4.2, Rule 15, p. 51
Classes and Normiq roadmaps
Rule 15 assigns: Class IIb, Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
Class IIb
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool