MDR · Annex VIII, Section 6.3 · ACTIVE DEVICES
MDR rule 11: Software
Updated on November 15, 2026
Official text
Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:
- death or an irreversible deterioration of a person's state of health, in which case it is in class III; or
- a serious deterioration of a person's state of health or a surgical intervention, in which case it is classified as class IIb.
Software intended to monitor physiological processes is classified as class IIa, except if it is intended for monitoring of vital physiological parameters, where the nature of variations of those parameters is such that it could result in immediate danger to the patient, in which case it is classified as class IIb.
All other software is classified as class I.
MDR, Annex VIII, Section 6.3, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/11
In plain words
Software that provides information used to take decisions with diagnostic or therapeutic purposes is in class IIa, class IIb if those decisions may cause a serious deterioration of a person’s state of health or a surgical intervention, and class III if they may cause death or an irreversible deterioration. Software intended to monitor physiological processes is in class IIa, or class IIb for vital parameters whose variations could result in immediate danger to the patient; all other software is in class I. Software that drives a device or influences its use takes that device’s class (Section 3.3).
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 4
‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.
- Software shall also be deemed to be an active device;
Examples from the MDCG guidance
The guidance lists these examples under rule 11 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class I
MDSW app intended to support conception by calculating the user’s fertility status based on a validated statistical algorithm.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 47
Class IIa
Cognitive therapy MDSW where a specialist determines the necessary cognitive therapy based on the outcome provided by the MDSW.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 46
MDSW intended to monitor physiological processes that are not considered to be vital.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 47
Class IIb
A mobile app intended to analyse a user’s heartbeat, detect abnormalities and inform a physician accordingly.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 46
Medical devices including MDSW intended to be used for continuous surveillance of vital physiological processes in anaesthesia, intensive care or emergency care.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 47
Class III
MDSW intended to perform diagnosis by means of image analysis for making treatment decisions in patients with acute stroke.
MDCG 2021-24 Rev.1, 4.2, Rule 11, p. 46
Classes and Normiq roadmaps
Rule 11 assigns: Class I, Class IIa, Class IIb, Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
Class I
Class IIa
Class IIb
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.