MDR · Annex VIII, Section 5.1 · INVASIVE DEVICES
MDR rule 5: Devices invasive with respect to body orifices
Updated on November 15, 2026
Official text
All invasive devices with respect to body orifices, other than surgically invasive devices, which are not intended for connection to an active device or which are intended for connection to a class I active device are classified as:
- class I if they are intended for transient use;
- class IIa if they are intended for short-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity, in which case they are classified as class I; and
- class IIb if they are intended for long-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity and are not liable to be absorbed by the mucous membrane, in which case they are classified as class IIa.
All invasive devices with respect to body orifices, other than surgically invasive devices, intended for connection to a class IIa, class IIb or class III active device, are classified as class IIa.
MDR, Annex VIII, Section 5.1, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/5
In plain words
The rule classifies devices invasive with respect to body orifices, other than surgically invasive devices, first by duration of use: class I for transient use, IIa for short-term and IIb for long-term use, one class lower in the oral cavity as far as the pharynx, an ear canal up to the ear drum or the nasal cavity (for long-term use, if they are not liable to be absorbed by the mucous membrane). Intended for connection to an active device of class IIa, IIb or III, the device is in class IIa.
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 6
‘invasive device’ means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body;
- MDR, Annex VIII, Section 2.1
‘Body orifice’ means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.
- MDR, Annex VIII, Section 2.2
‘Surgically invasive device’ means:
- an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and
- a device which produces penetration other than through a body orifice.
- MDR, Article 2, point 5
‘implantable device’ means any device, including those that are partially or wholly absorbed, which is intended:
- to be totally introduced into the human body, or
- to replace an epithelial surface or the surface of the eye,
- by clinical intervention and which is intended to remain in place after the procedure.
- Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device;
- MDR, Annex VIII, Section 1.1
‘Transient’ means normally intended for continuous use for less than 60 minutes.
- MDR, Annex VIII, Section 1.2
‘Short term’ means normally intended for continuous use for between 60 minutes and 30 days.
- MDR, Annex VIII, Section 1.3
‘Long term’ means normally intended for continuous use for more than 30 days.
- MDR, Annex VIII, Section 3.6
In calculating the duration referred to in Section 1, continuous use shall mean:
- the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and
- the accumulated use of a device that is intended by the manufacturer to be replaced immediately with another of the same type.
Examples from the MDCG guidance
The guidance lists these examples under rule 5 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class I
Examination gloves
MDCG 2021-24 Rev.1, 4.2, Rule 5, p. 34
Class IIa
Tracheal tubes
MDCG 2021-24 Rev.1, 4.2, Rule 5, p. 34
Tracheostomy or tracheal tubes connected to a ventilator
MDCG 2021-24 Rev.1, 4.2, Rule 5, p. 35
Class IIb
Long term corrective contact lenses
MDCG 2021-24 Rev.1, 4.2, Rule 5, p. 34
Regarding devices intended for connection to an active device: the strictest rule and sub-rule resulting in higher classification will apply. For instance a trachea cannula for long-term use need to be classified as class IIb.
MDCG 2021-24 Rev.1, 4.2, Rule 5, Practical issues of classification, p. 35
Classes and Normiq roadmaps
Rule 5 assigns: Class I, Class IIa, Class IIb. For each, the Normiq roadmap on every market where Annex VIII applies.
Class I
Class IIa
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.