IVDR · Annex VIII, Section 2.5 · CLASSIFICATION RULES
IVDR rule 5: Laboratory products, instruments and specimen receptacles
Updated on November 15, 2026
Official text
The following devices are classified as class A:
- products for general laboratory use, accessories which possess no critical characteristics, buffer solutions, washing solutions, and general culture media and histological stains, intended by the manufacturer to make them suitable for in vitro diagnostic procedures relating to a specific examination;
- instruments intended by the manufacturer specifically to be used for in vitro diagnostic procedures;
- specimen receptacles.
IVDR, Annex VIII, Section 2.5, consolidated version of January 10, 2025.
Short link: normiq.app/en/ivdr/annexe/viii/5
In plain words
Class A covers products for general laboratory use, accessories which possess no critical characteristics, buffer solutions, washing solutions, general culture media and histological stains intended by the manufacturer for in vitro diagnostic procedures relating to a specific examination; instruments intended by the manufacturer specifically for in vitro diagnostic procedures; and specimen receptacles.
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 5 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).
Class A
Solutions like cleaners, buffer solutions, lysing solutions, diluents specified for use with an IVD.
MDCG 2020-16 rev.5, 4, Rule 5(a), p. 40
Enzyme immunoassay analyser, PCR thermocycler, sequencer for NGS applications, clinical chemistry analyser.
MDCG 2020-16 rev.5, 4, Rule 5(b), p. 41
A standalone urine collection cup, a stool container, a saliva collection tube or a blood spot collection card (e.g. specimen collection via finger-prick) intended for subsequent in vitro diagnostic examination are in class A.
MDCG 2020-16 rev.5, 4, Rule 5(c), p. 43
Blood gas analyser and the software driving and influencing it are classified as either: Class A (Rule 5b) where the analyser does not directly measure a parameter from a specimen, rather the specimen is analysed using replaceable sensor cassette / reagents.
MDCG 2020-16 rev.5, Annex 1, Example 2, p. 48
Classes and Normiq roadmaps
Rule 5 assigns: Class A. For each, the Normiq roadmap on every market where Annex VIII applies.
The same rule in Switzerland
Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.