MDR · Annex VIII, Section 4.3 · NON-INVASIVE DEVICES
MDR rule 3: Modifying the composition of blood, tissues or liquids
Updated on November 15, 2026
Official text
All non-invasive devices intended for modifying the biological or chemical composition of human tissues or cells, blood, other body liquids or other liquids intended for implantation or administration into the body are classified as class IIb, unless the treatment for which the device is used consists of filtration, centrifugation or exchanges of gas, heat, in which case they are classified as class IIa.
All non-invasive devices consisting of a substance or a mixture of substances intended to be used in vitro in direct contact with human cells, tissues or organs taken from the human body or used in vitro with human embryos before their implantation or administration into the body are classified as class III.
MDR, Annex VIII, Section 4.3, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/3
In plain words
A non-invasive device that modifies the biological or chemical composition of human tissues or cells, blood or liquids intended for implantation or administration is in class IIb, or class IIa where the treatment consists of filtration, centrifugation or exchanges of gas or heat. A substance or mixture of substances used in vitro in direct contact with human cells, tissues or organs taken from the body, or with human embryos before their implantation or administration, is in class III.
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 3 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class IIa
Warming or cooling the blood in an extracorporeal circulation system.
MDCG 2021-24 Rev.1, 4.2, Rule 3, p. 31
Class IIb
Devices intended to remove undesirable substances out of the blood by exchange of solutes such as hemodialysers
MDCG 2021-24 Rev.1, 4.2, Rule 3, p. 31
Class III
IVF cell media without human albumin
MDCG 2021-24 Rev.1, 4.2, Rule 3, p. 32
Classes and Normiq roadmaps
Rule 3 assigns: Class IIa, Class IIb, Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
Class IIa
Class IIb
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool