MDR · Annex VIII, Section 6.1 · ACTIVE DEVICES

MDR rule 9: Active therapeutic devices

Updated on November 15, 2026

Official text

All active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb.

All active devices intended to control or monitor the performance of active therapeutic class IIb devices, or intended directly to influence the performance of such devices are classified as class IIb.

All active devices intended to emit ionizing radiation for therapeutic purposes, including devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.

All active devices that are intended for controlling, monitoring or directly influencing the performance of active implantable devices are classified as class III.

MDR, Annex VIII, Section 6.1, consolidated version of July 19, 2026.

Short link: normiq.app/en/mdr/annexe/viii/9

In plain words

An active therapeutic device intended to administer or exchange energy is in class IIa, or class IIb if it may do so in a potentially hazardous way, taking account of the nature, density and site of application of the energy. Active devices that control, monitor or directly influence an active therapeutic class IIb device, and those that emit ionizing radiation for therapeutic purposes, are also in class IIb; those that control, monitor or directly influence an active implantable device are in class III.

Explanation written by Normiq; only the official text above is authoritative.

Definitions the rule relies on

MDR, Article 2, point 4

‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.

  • Software shall also be deemed to be an active device;
MDR, Annex VIII, Section 2.4

‘Active therapeutic device’ means any active device used, whether alone or in combination with other devices, to support, modify, replace or restore biological functions or structures with a view to treatment or alleviation of an illness, injury or disability.

Examples from the MDCG guidance

The guidance lists these examples under rule 9 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).

Class IIa

  • TENS devices

    MDCG 2021-24 Rev.1, 4.2, Rule 9, p. 42

  • external hearing aids

    MDCG 2021-24 Rev.1, 4.2, Rule 9, p. 42

Class IIb

  • lung ventilators

    MDCG 2021-24 Rev.1, 4.2, Rule 9, p. 42

  • surgical lasers

    MDCG 2021-24 Rev.1, 4.2, Rule 9, p. 43

Class III

  • Remote monitoring devices for active implantable devices

    MDCG 2021-24 Rev.1, 4.2, Rule 9, p. 43

Classes and Normiq roadmaps

Rule 9 assigns: Class IIa, Class IIb, Class III. For each, the Normiq roadmap on every market where Annex VIII applies.

The same rule in Switzerland

Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

Art. 15 MedDO

The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained