MDR · Annex VIII, Section 7.8 · SPECIAL RULES
MDR rule 21: Devices composed of substances
Updated on November 15, 2026
Official text
Devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body are classified as:
- class III if they, or their products of metabolism, are systemically absorbed by the human body in order to achieve the intended purpose;
- class III if they achieve their intended purpose in the stomach or lower gastrointestinal tract and they, or their products of metabolism, are systemically absorbed by the human body;
- class IIa if they are applied to the skin or if they are applied in the nasal or oral cavity as far as the pharynx, and achieve their intended purpose on those cavities; and
- class IIb in all other cases.
MDR, Annex VIII, Section 7.8, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/21
In plain words
A device composed of substances introduced via a body orifice or applied to the skin, and absorbed by or locally dispersed in the body, is in class III if it or its products of metabolism are systemically absorbed to achieve its intended purpose, or if it achieves its purpose in the stomach or lower gastrointestinal tract and it or its products of metabolism are systemically absorbed; in class IIa if applied to the skin or in the nasal or oral cavity as far as the pharynx and achieving its purpose there; in class IIb in all other cases.
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 6
‘invasive device’ means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body;
- MDR, Annex VIII, Section 2.1
‘Body orifice’ means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.
- MDR, Annex VIII, Section 2.2
‘Surgically invasive device’ means:
- an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and
- a device which produces penetration other than through a body orifice.
- MDR, Article 2, point 5
‘implantable device’ means any device, including those that are partially or wholly absorbed, which is intended:
- to be totally introduced into the human body, or
- to replace an epithelial surface or the surface of the eye,
- by clinical intervention and which is intended to remain in place after the procedure.
- Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device;
Examples from the MDCG guidance
The guidance lists these examples under rule 21 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class IIa
Salt water used e.g. as nose or throat sprays
MDCG 2021-24 Rev.1, 4.2, Rule 21, p. 58
Class IIb
Simethicone preparations for oral administration
MDCG 2021-24 Rev.1, 4.2, Rule 21, p. 58
Eye drops for hydration
MDCG 2021-24 Rev.1, 4.2, Rule 21, p. 58
Classes and Normiq roadmaps
Rule 21 assigns: Class IIa, Class IIb, Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
Class IIa
Class IIb
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained