MDR · Annex VIII, Section 7.1 · SPECIAL RULES
MDR rule 14: Devices incorporating a medicinal substance
Updated on November 15, 2026
Official text
All devices incorporating, as an integral part, a substance which, if used separately, can be considered to be a medicinal product, as defined in point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma, as defined in point 10 of Article 1 of that Directive, and that has an action ancillary to that of the devices, are classified as class III.
MDR, Annex VIII, Section 7.1, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/14
In plain words
A device that incorporates, as an integral part, a substance which, if used separately, can be considered a medicinal product within the meaning of Directive 2001/83/EC, including one derived from human blood or plasma, with an action ancillary to that of the device, is in class III.
Explanation written by Normiq; only the official text above is authoritative.
Examples from the MDCG guidance
The guidance lists these examples under rule 14 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class III
Bone cement with antibiotics
MDCG 2021-24 Rev.1, 4.2, Rule 14, p. 50
Drug eluting stents (e.g. coronary, pulmonary)
MDCG 2021-24 Rev.1, 4.2, Rule 14, p. 50
Catheters coated with anticoagulants (e. g. heparin)
MDCG 2021-24 Rev.1, 4.2, Rule 14, p. 50
Classes and Normiq roadmaps
Rule 14 assigns: Class III. For each, the Normiq roadmap on every market where Annex VIII applies.
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool