MDR · Annex VIII, Section 6.4 · ACTIVE DEVICES
MDR rule 12: Administering or removing medicinal products, body liquids or substances
Updated on November 15, 2026
Official text
All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are classified as class IIa, unless this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application in which case they are classified as class IIb.
MDR, Annex VIII, Section 6.4, consolidated version of July 19, 2026.
Short link: normiq.app/en/mdr/annexe/viii/12
In plain words
An active device that administers and/or removes medicinal products, body liquids or other substances to or from the body is in class IIa, or class IIb if it does so in a potentially hazardous manner, taking account of the nature of the substances, the part of the body concerned and the mode of application.
Explanation written by Normiq; only the official text above is authoritative.
Definitions the rule relies on
- MDR, Article 2, point 4
‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.
- Software shall also be deemed to be an active device;
Examples from the MDCG guidance
The guidance lists these examples under rule 12 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).
Class IIa
Feeding pumps
MDCG 2021-24 Rev.1, 4.2, Rule 12, p. 48
Class IIb
Infusion pumps
MDCG 2021-24 Rev.1, 4.2, Rule 12, p. 48
Anaesthesia machines
MDCG 2021-24 Rev.1, 4.2, Rule 12, p. 48
Classes and Normiq roadmaps
Rule 12 assigns: Class IIa, Class IIb. For each, the Normiq roadmap on every market where Annex VIII applies.
Class IIa
The same rule in Switzerland
Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.
Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.
The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: MDR classification: the 22 rules of Annex VIII explained
Classify your device
The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.
Open the classification tool