IVDR · Annex VIII, Section 2.7 · CLASSIFICATION RULES

IVDR rule 7: Controls without an assigned value

Updated on November 15, 2026

Official text

Devices which are controls without a quantitative or qualitative assigned value are classified as class B.

IVDR, Annex VIII, Section 2.7, consolidated version of January 10, 2025.

Short link: normiq.app/en/ivdr/annexe/viii/7

In plain words

A control without a quantitative or qualitative assigned value is in class B. A control material with quantitative or qualitative assigned values intended for one or more analytes instead takes the class of the device (Section 1.6).

Explanation written by Normiq; only the official text above is authoritative.

Examples from the MDCG guidance

The guidance lists these examples under rule 7 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).

Class B

  • Unassigned control sera.

    MDCG 2020-16 rev.5, 4, Rule 7, p. 45

  • Non-assay specific control plasma for use in coagulation.

    MDCG 2020-16 rev.5, 4, Rule 7, p. 45

Classes and Normiq roadmaps

Rule 7 assigns: Class B. For each, the Normiq roadmap on every market where Annex VIII applies.

The same rule in Switzerland

Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

Art. 14 IvDO