IVDR · Annex VIII, Section 2.1 · CLASSIFICATION RULES

IVDR rule 1: Transmissible agents: donations, propagation, infectious load

Updated on November 15, 2026

Official text

Devices intended to be used for the following purposes are classified as class D:

  • detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration;
  • detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation;
  • determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.

IVDR, Annex VIII, Section 2.1, consolidated version of January 10, 2025.

Short link: normiq.app/en/ivdr/annexe/viii/1

In plain words

Class D covers devices intended to detect the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or their derivatives, to assess their suitability for transfusion, transplantation or cell administration; to detect the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; or to determine the infectious load of a life-threatening disease where monitoring is critical in patient management.

Explanation written by Normiq; only the official text above is authoritative.

Examples from the MDCG guidance

The guidance lists these examples under rule 1 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).

Class D

  • A device intended to screen blood and tissue donations for syphilis would fall under class D according to rule 1.

    MDCG 2020-16 rev.5, 3.1 Intended purpose, p. 5

  • Devices intended to be used for determining the infectious load of: Human Immunodeficiency Virus.

    MDCG 2020-16 rev.5, 4, Rule 1 third indent, p. 13

Classes and Normiq roadmaps

Rule 1 assigns: Class D. For each, the Normiq roadmap on every market where Annex VIII applies.

The same rule in Switzerland

Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

Art. 14 IvDO