title: "Swissmedic vigilance"
url: https://normiq.app/en/regulations/ch-vigilance
language: en
updated: 2026-09-03
---

# Swissmedic vigilance

CH-VIGILANCE

Reporting of incidents and recalls to Swissmedic

- Category

  Post-market

- Market

  Switzerland

- Applicable classes

  Class I medical device, Class IIa medical device, Class IIb medical device, Class III medical device

## In practice

MedDO Article 66 adopts the MDR deadlines: 15 days as a general rule, 10 days in the event of death or an unanticipated serious deterioration in health, 2 days in the event of a serious public health threat, counted from the day the manufacturer becomes aware of the incident. Field safety corrective actions are also notified to Swissmedic. The periodic safety update report falls under MedDO Articles 60 to 62, not vigilance.

## Official sources

<https://www.fedlex.admin.ch/eli/cc/2020/552/fr>