# MDR rule 17: Recording X-ray diagnostic images

MDR · Annex VIII, Section 7.4 · SPECIAL RULES

Updated on November 15, 2026

## Official text

Devices specifically intended for recording of diagnostic images generated by X-ray radiation are classified as class IIa.

MDR, Annex VIII, Section 7.4, consolidated version of July 19, 2026.

Short link: [normiq.app/en/mdr/annexe/viii/17](https://normiq.app/en/mdr/annexe/viii/17)

## In plain words

A device specifically intended for recording diagnostic images generated by X-ray radiation is in class IIa.

Explanation written by Normiq; only the official text above is authoritative.

## Examples from the MDCG guidance

The guidance lists these examples under rule 17 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).

### Class IIa

- Digital x-ray detectors for recording images

  MDCG 2021-24 Rev.1, 4.2, Rule 17, p. 53

- X-ray films

  MDCG 2021-24 Rev.1, 4.2, Rule 17, p. 53

## Classes and Normiq roadmaps

Rule 17 assigns: Class IIa. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class IIa

- [European Union 15 phases · 66 steps · estimated duration 240 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-iia)
- [Switzerland 16 phases · 74 steps · estimated duration 120 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-iia)
- [European Union and Switzerland 17 phases · 82 steps · estimated duration 300 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-iia)

## The same rule in Switzerland

> Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 15 MedDO](https://www.fedlex.admin.ch/eli/cc/2020/552/en#art_15)

The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: [MDR classification: the 22 rules of Annex VIII explained](https://normiq.app/en/guides/classification-mdr-annexe-viii)

## Classify your device

The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [MDR, consolidated version of July 19, 2026 (02017R0745-20260719), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20260719)
- [MDCG 2021-24 Rev.1, “Guidance on classification of medical devices”, April 2026, p. 53](https://health.ec.europa.eu/document/download/cbb19821-a517-4e13-bf87-fdc6ddd1782e_en?filename=mdcg_2021-24_en.pdf)