title: "MDR rule 16: Disinfection, sterilisation and contact lens care"
url: https://normiq.app/en/classification/mdr/rule-16
language: en
updated: 2026-11-15
---

# MDR rule 16: Disinfection, sterilisation and contact lens care

MDR · Annex VIII, Section 7.3 · SPECIAL RULES

Updated on November 15, 2026

## Official text

All devices intended specifically to be used for disinfecting, cleaning, rinsing or, where appropriate, hydrating contact lenses are classified as class IIb.

All devices intended specifically to be used for disinfecting or sterilising medical devices are classified as class IIa, unless they are disinfecting solutions or washer-disinfectors intended specifically to be used for disinfecting invasive devices, as the end point of processing, in which case they are classified as class IIb.

This rule does not apply to devices that are intended to clean devices other than contact lenses by means of physical action only.

MDR, Annex VIII, Section 7.3, consolidated version of July 19, 2026.

Short link: [normiq.app/en/mdr/annexe/viii/16](https://normiq.app/en/mdr/annexe/viii/16)

## In plain words

Devices intended specifically for disinfecting, cleaning, rinsing or, where appropriate, hydrating contact lenses are in class IIb. Those for disinfecting or sterilising medical devices are in class IIa, except disinfecting solutions and washer-disinfectors intended for disinfecting invasive devices as the end point of processing, in class IIb. The rule does not cover cleaning by physical action only, other than of contact lenses.

Explanation written by Normiq; only the official text above is authoritative.

## Examples from the MDCG guidance

The guidance lists these examples under rule 16 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).

### Class IIa

- Sterilisers intended to sterilise medical devices (invasive or non-invasive) in a medical environment

  MDCG 2021-24 Rev.1, 4.2, Rule 16, p. 52

### Class IIb

- Contact lens storing solutions

  MDCG 2021-24 Rev.1, 4.2, Rule 16, p. 52

- Washer-disinfector equipment specifically for disinfecting endoscopes or other invasive devices at the end point of processing (e. g. dental equipment)

  MDCG 2021-24 Rev.1, 4.2, Rule 16, p. 52

## Classes and Normiq roadmaps

Rule 16 assigns: Class IIa, Class IIb. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class IIa

- [European Union 15 phases · 66 steps · estimated duration 240 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-iia)
- [Switzerland 16 phases · 74 steps · estimated duration 120 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-iia)
- [European Union and Switzerland 17 phases · 82 steps · estimated duration 300 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-iia)

### Class IIb

- [European Union 15 phases · 68 steps · estimated duration 360 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-iib)
- [Switzerland 16 phases · 77 steps · estimated duration 180 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-iib)
- [European Union and Switzerland 17 phases · 85 steps · estimated duration 420 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-iib)

## The same rule in Switzerland

> Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 15 MedDO](https://www.fedlex.admin.ch/eli/cc/2020/552/en#art_15)

The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: [MDR classification: the 22 rules of Annex VIII explained](https://normiq.app/en/guides/classification-mdr-annexe-viii)

## Classify your device

The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [MDR, consolidated version of July 19, 2026 (02017R0745-20260719), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20260719)
- [MDCG 2021-24 Rev.1, “Guidance on classification of medical devices”, April 2026, p. 52](https://health.ec.europa.eu/document/download/cbb19821-a517-4e13-bf87-fdc6ddd1782e_en?filename=mdcg_2021-24_en.pdf)