title: "MDR rule 10: Active devices for diagnosis and monitoring"
url: https://normiq.app/en/classification/mdr/rule-10
language: en
updated: 2026-11-15
---

# MDR rule 10: Active devices for diagnosis and monitoring

MDR · Annex VIII, Section 6.2 · ACTIVE DEVICES

Updated on November 15, 2026

## Official text

Active devices intended for diagnosis and monitoring are classified as class IIa:

- if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient's body, in the visible spectrum, in which case they are classified as class I;
- if they are intended to image in vivo distribution of radiopharmaceuticals; or
- if they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or they are intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb.

Active devices intended to emit ionizing radiation and intended for diagnostic or therapeutic radiology, including interventional radiology devices and devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.

MDR, Annex VIII, Section 6.2, consolidated version of July 19, 2026.

Short link: [normiq.app/en/mdr/annexe/viii/10](https://normiq.app/en/mdr/annexe/viii/10)

## In plain words

An active device for diagnosis or monitoring is in class IIa if it supplies energy absorbed by the body (class I if it illuminates the body in the visible spectrum), if it images the in vivo distribution of radiopharmaceuticals, or if it allows direct diagnosis or monitoring of vital physiological processes. It is in class IIb if it monitors vital parameters whose variations could result in immediate danger to the patient, or is used for diagnosis where the patient is in immediate danger; devices emitting ionizing radiation for diagnostic or therapeutic radiology are also in class IIb.

Explanation written by Normiq; only the official text above is authoritative.

## Definitions the rule relies on

- MDR, Article 2, point 4

  ‘active device’ means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.

  - Software shall also be deemed to be an active device;

- MDR, Annex VIII, Section 2.5

  ‘Active device intended for diagnosis and monitoring’ means any active device used, whether alone or in combination with other devices, to supply information for detecting, diagnosing, monitoring or treating physiological conditions, states of health, illnesses or congenital deformities.

- MDR, Annex VIII, Section 3.7

  A device is considered to allow direct diagnosis when it provides the diagnosis of the disease or condition in question by itself or when it provides decisive information for the diagnosis.

## Examples from the MDCG guidance

The guidance lists these examples under rule 10 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).

### Class I

- Examination lamps

  MDCG 2021-24 Rev.1, 4.2, Rule 10, p. 44

### Class IIa

- Diagnostic ultrasound

  MDCG 2021-24 Rev.1, 4.2, Rule 10, p. 44

- Electrocardiographs

  MDCG 2021-24 Rev.1, 4.2, Rule 10, p. 44

### Class IIb

- Apnoea monitors, including apnoea monitors in home care

  MDCG 2021-24 Rev.1, 4.2, Rule 10, p. 44

- Diagnostic X-Ray machine

  MDCG 2021-24 Rev.1, 4.2, Rule 10, p. 45

## Classes and Normiq roadmaps

Rule 10 assigns: Class I, Class IIa, Class IIb. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class I

- [European Union 14 phases · 63 steps · estimated duration 210 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-i)
- [Switzerland 14 phases · 69 steps · estimated duration 60 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-i)
- [European Union and Switzerland 15 phases · 77 steps · estimated duration 240 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-i)

### Class IIa

- [European Union 15 phases · 66 steps · estimated duration 240 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-iia)
- [Switzerland 16 phases · 74 steps · estimated duration 120 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-iia)
- [European Union and Switzerland 17 phases · 82 steps · estimated duration 300 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-iia)

### Class IIb

- [European Union 15 phases · 68 steps · estimated duration 360 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-iib)
- [Switzerland 16 phases · 77 steps · estimated duration 180 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-iib)
- [European Union and Switzerland 17 phases · 85 steps · estimated duration 420 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-iib)

## The same rule in Switzerland

> Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 15 MedDO](https://www.fedlex.admin.ch/eli/cc/2020/552/en#art_15)

The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: [MDR classification: the 22 rules of Annex VIII explained](https://normiq.app/en/guides/classification-mdr-annexe-viii)

## Classify your device

The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [MDR, consolidated version of July 19, 2026 (02017R0745-20260719), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20260719)
- [MDCG 2021-24 Rev.1, “Guidance on classification of medical devices”, April 2026, p. 44, 45](https://health.ec.europa.eu/document/download/cbb19821-a517-4e13-bf87-fdc6ddd1782e_en?filename=mdcg_2021-24_en.pdf)