MDR · Annex VIII, Section 4.1 · NON-INVASIVE DEVICES

Updated on November 15, 2026

## Official text

All non-invasive devices are classified as class I, unless one of the rules set out hereinafter applies.

MDR, Annex VIII, Section 4.1, consolidated version of July 19, 2026.

Short link: [normiq.app/en/mdr/annexe/viii/1](https://normiq.app/en/mdr/annexe/viii/1)

## In plain words

Every non-invasive device is in class I, unless one of the rules that follow applies. It therefore applies only by default: as soon as another rule covers the same device, the strictest one gives the class (Section 3.5).

Explanation written by Normiq; only the official text above is authoritative.

## Definitions the rule relies on

- MDR, Article 2, point 6

  ‘invasive device’ means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body;

- MDR, Annex VIII, Section 2.1

  ‘Body orifice’ means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.

- MDR, Annex VIII, Section 2.2

  ‘Surgically invasive device’ means:

  - an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and
  - a device which produces penetration other than through a body orifice.

- MDR, Article 2, point 5

  ‘implantable device’ means any device, including those that are partially or wholly absorbed, which is intended:

  - to be totally introduced into the human body, or
  - to replace an epithelial surface or the surface of the eye,
  - by clinical intervention and which is intended to remain in place after the procedure.
  - Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device;

## Examples from the MDCG guidance

The guidance lists these examples under rule 1 (MDCG 2021-24 Rev.1). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 3.1).

### Class I

- Stethoscopes

  MDCG 2021-24 Rev.1, 4.2, Rule 1, p. 28

- Corrective spectacle frames (i.e. glasses) and lenses in frames

  MDCG 2021-24 Rev.1, 4.2, Rule 1, p. 28

- Wheelchairs pushed by hand

  MDCG 2021-24 Rev.1, 4.2, Rule 1, p. 28

## Classes and Normiq roadmaps

Rule 1 assigns: Class I. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class I

- [European Union 14 phases · 63 steps · estimated duration 210 days See the roadmap →](https://normiq.app/en/roadmap/eu/class-i)
- [Switzerland 14 phases · 69 steps · estimated duration 60 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/class-i)
- [European Union and Switzerland 15 phases · 77 steps · estimated duration 240 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/class-i)

## The same rule in Switzerland

> Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-MDR, taking account of the implementing acts of the European Commission listed in Annex 5a.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 15 MedDO](https://www.fedlex.admin.ch/eli/cc/2020/552/en#art_15)

The families of rules, the cross-cutting criteria, Rule 11 and common pitfalls are explained in the guide: [MDR classification: the 22 rules of Annex VIII explained](https://normiq.app/en/guides/classification-mdr-annexe-viii)

## Classify your device

The tool asks the questions of the 22 rules of the MDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [MDR, consolidated version of July 19, 2026 (02017R0745-20260719), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20260719)
- [MDCG 2021-24 Rev.1, “Guidance on classification of medical devices”, April 2026, p. 28](https://health.ec.europa.eu/document/download/cbb19821-a517-4e13-bf87-fdc6ddd1782e_en?filename=mdcg_2021-24_en.pdf)