# IVDR rule 7: Controls without an assigned value

IVDR · Annex VIII, Section 2.7 · CLASSIFICATION RULES

Updated on November 15, 2026

## Official text

Devices which are controls without a quantitative or qualitative assigned value are classified as class B.

IVDR, Annex VIII, Section 2.7, consolidated version of January 10, 2025.

Short link: [normiq.app/en/ivdr/annexe/viii/7](https://normiq.app/en/ivdr/annexe/viii/7)

## In plain words

A control without a quantitative or qualitative assigned value is in class B. A control material with quantitative or qualitative assigned values intended for one or more analytes instead takes the class of the device (Section 1.6).

Explanation written by Normiq; only the official text above is authoritative.

## Examples from the MDCG guidance

The guidance lists these examples under rule 7 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).

### Class B

- Unassigned control sera.

  MDCG 2020-16 rev.5, 4, Rule 7, p. 45

- Non-assay specific control plasma for use in coagulation.

  MDCG 2020-16 rev.5, 4, Rule 7, p. 45

## Classes and Normiq roadmaps

Rule 7 assigns: Class B. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class B

- [European Union 15 phases · 61 steps · estimated duration 180 days See the roadmap →](https://normiq.app/en/roadmap/eu/ivd-class-b)
- [Switzerland 16 phases · 59 steps · estimated duration 90 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/ivd-class-b)
- [European Union and Switzerland 16 phases · 64 steps · estimated duration 225 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/ivd-class-b)

## The same rule in Switzerland

> Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 14 IvDO](https://www.fedlex.admin.ch/eli/cc/2022/291/en#art_14)

## Classify your device

The tool asks the questions of the 7 rules of the IVDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [IVDR, consolidated version of January 10, 2025 (02017R0746-20250110), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0746-20250110)
- [MDCG 2020-16 rev.5, “Guidance on Classification Rules for In Vitro Diagnostic Medical Devices under Regulation (EU) 2017/746”, September 2026, p. 45](https://health.ec.europa.eu/document/download/12f9756a-1e0d-4aed-9783-d948553f1705_en?filename=md_mdcg_2020_guidance_classification_ivd-md_en.pdf)