title: "IVDR rule 1: Transmissible agents: donations, propagation, infectious load"
url: https://normiq.app/en/classification/ivdr/rule-1
language: en
updated: 2026-11-15
---

# IVDR rule 1: Transmissible agents: donations, propagation, infectious load

IVDR · Annex VIII, Section 2.1 · CLASSIFICATION RULES

Updated on November 15, 2026

## Official text

Devices intended to be used for the following purposes are classified as class D:

- detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration;
- detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation;
- determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.

IVDR, Annex VIII, Section 2.1, consolidated version of January 10, 2025.

Short link: [normiq.app/en/ivdr/annexe/viii/1](https://normiq.app/en/ivdr/annexe/viii/1)

## In plain words

Class D covers devices intended to detect the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or their derivatives, to assess their suitability for transfusion, transplantation or cell administration; to detect the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; or to determine the infectious load of a life-threatening disease where monitoring is critical in patient management.

Explanation written by Normiq; only the official text above is authoritative.

## Examples from the MDCG guidance

The guidance lists these examples under rule 1 (MDCG 2020-16 rev.5). It is not legally binding, and the class always depends on the intended purpose of the device (Annex VIII, Section 1.1).

### Class D

- A device intended to screen blood and tissue donations for syphilis would fall under class D according to rule 1.

  MDCG 2020-16 rev.5, 3.1 Intended purpose, p. 5

- Devices intended to be used for determining the infectious load of: Human Immunodeficiency Virus.

  MDCG 2020-16 rev.5, 4, Rule 1 third indent, p. 13

## Classes and Normiq roadmaps

Rule 1 assigns: Class D. For each, the Normiq roadmap on every market where Annex VIII applies.

### Class D

- [European Union 15 phases · 65 steps · estimated duration 360 days See the roadmap →](https://normiq.app/en/roadmap/eu/ivd-class-d)
- [Switzerland 16 phases · 63 steps · estimated duration 180 days See the roadmap →](https://normiq.app/en/roadmap/switzerland/ivd-class-d)
- [European Union and Switzerland 16 phases · 68 steps · estimated duration 450 days See the roadmap →](https://normiq.app/en/roadmap/eu-switzerland/ivd-class-d)

## The same rule in Switzerland

> Devices shall be divided into classes A, B, C and D, taking into account the intended purpose of the devices and their inherent risks. This classification must comply with the provisions of Annex VIII to EU-IVDR.

Fedlex’s English translation, which has no legal force: the ordinance is official only in French, German and Italian.

[Art. 14 IvDO](https://www.fedlex.admin.ch/eli/cc/2022/291/en#art_14)

## Classify your device

The tool asks the questions of the 7 rules of the IVDR, one at a time, and gives the indicative class with the rule and its text.

[Open the classification tool](https://normiq.app/en/classification#outil)

## Official sources

- [IVDR, consolidated version of January 10, 2025 (02017R0746-20250110), EUR-Lex](https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0746-20250110)
- [MDCG 2020-16 rev.5, “Guidance on Classification Rules for In Vitro Diagnostic Medical Devices under Regulation (EU) 2017/746”, September 2026, p. 5, 13](https://health.ec.europa.eu/document/download/12f9756a-1e0d-4aed-9783-d948553f1705_en?filename=md_mdcg_2020_guidance_classification_ivd-md_en.pdf)